Checking the assemblies 2 - 5
Siemens AG SPR8-125.061.02 Page 5 of 8 POLYMOBIL Plus
Medical Engineering Rev. 01 01.00 TD SD 24
Reset - switch
SW2
D925 Rotating anode starter, filament inverter, ON/OFF switching circuit 2
LED:
Fuses:
VP +15 V for inverter frequency ok
+12V +12.5 V auxiliary voltage ok
I/O +12.5 V auxiliary voltage for I/O ok
AR Line voltage for anode starter ok
+DC, -DC 300 V voltage for filament ok
F1 10 AT + 12.5V for collimator lamp
F3 0.6 AT Auxiliary voltage +12.5 V for I/O
F4 0.6 AT Auxiliary voltage +12.5 V
F5 3.2 AF Line voltage for anode starter
F6 1.0 AT Line voltage for filament circuit
F7 10 AT Power
F8 10 AT Power
Fig. 6
ST
P1
P2
P3
P4
P5
P6
X11
X10 X20
X12
P7
GND JR
REG CAL KVP
KV KVN IH JRS KVS F1 I I1 I2
TRG JIM SCH GND
VC BAT
KVM
SCM VR
Service
V22
V23 V24 V25
1
1 1
X15
SW2
D 915
V34