3 - 1
Siemens AG SPR8-125.061.02 Page 1 of 4 POLYMOBIL Plus
Medical Engineering Rev. 01 01.00 TD SD 24
Troubleshooting 3
Error messages 3
a) During initialization
b)During stand by
c)During exposure
Error messages Err 1, and initialization errors Err 90, Err 96 and
Err 97 block the generator so that the system must be switched
off and switched back on again.
"StandBy" error messages disappear when the error is cleared.
Error messages which occur during an exposure will continue
flashing until another key is activated.
Err 90 EPROM Checksum
Err 96 KVnom error
Err 97 mAnom error
Err 1 Capacity charge error Charging circuit error
Err 2 +15 VError
Err 3 Vfil < Vstby
Err 4 Vfil > Vstby
Err 5 KVact 0
Err 6 mAact 0
Err 7 Rot 0
Err 8 Braking failure Error in the rotating anode circuit
Err 9 Pressure switch Oil pressure switch in the single tank
is active
Err 30 Fast capacitor charge Capacitor bank is charging too fast
Err 31 Slow capacitor charge Capacitor bank is charging too slow
Err 33 Inverter Error
Err 34 Filament circuit Inverter Short-Cir-
cuit
Err 10 Rot
= 0 No rotating anode acceleration
Err 11 Short circuit in main inverter Short circuit in the main inverter
Err 12 KVmax
Err 13 Imax
Err 14 KVact < KVnom
Err 15 mAact < mAnom
Err 17 Maximum exposure time Max. exposure reached
Err 18 Exposure not completed Exposure time not completed(expo-
sure release switch released)
Err 20 Capacitor not ready Capacitor bank not completely
charged
Err 21 Vfil > Vmax during prep Maximum filament heating exceeded
during preparation
Err 22 Maximum prep time Max. preparation time exceeded