POLYMOBIL Plus SPR8-125.061.02 Page 8 of 8 Siemens AG
Rev. 01 01.00 TD SD 24 Medical Engineering
2 - 8 Checking the assemblies
Status of the LED’s (optimum condition) 2
a) POLYMOBIL Plus OFF, line voltage plug connected:
b) POLYMOBIL Plus ON
, standby
c) Preparation
All LED’s as described under b) and also V22 on D915.
d) Exposure
All LED’s as described under b) and also V22 and V23 as well as radiation indicator V34
on D915.
In case of an error, V24 on D915 glow (red).
D925: LED +12 V ON
LED I/O ON
D950: LED 450 Vcc (V30) glows for 10 minutes following shutdown
D970: all LED’s glows for 10 minutes following shutdown
D915: The exposure values are displayed
D925: LED VP ON
LED +12 V ON
LED I/O ON
LED AR ON
LED +DC ON
LED -DC ON
D950: LED LINE (V25) ON
LED 450 Vcc (V30) ON
D970: all LED’s ON