Siemens AG SPR8-125.061.02 Page 1 of 2 POLYMOBIL Plus
Medical Engineering Rev. 01 01.00 TD SD 24
Service programs 4
To bypass the service mode, the ST slot (1/Fig.1, below the display on D915) has to be
jumpered before switching on the POLYMOBIL Plus.
As soon as "Pr. 1" appears in the display, bring the jumper back to the "OPEN" position.
To quit the service mode, switch the POLYMOBIL Plus off/on or press the Reset button
SW 2 (2/Fig.1).
KV +/- keys toggle between the programs. When you are
already in a program, the next or previous program step
is selected.
Setting of the kV values in "Pr. 5" and "Pr. 6".
mAs +/- Setting of the mAs values in "Pr. 5" and "Pr. 6".
Lamp switch Activating or quitting the current program