Alarm, error, and system messages
9.1 Diagnostics
SITRANS F M MAG 5000/6000
56 Operating Instructions, 01/2010, SFIDK.PS.026.G1.02
3HQGLQJ
(UURUWH[W
Status log
The latest 9 errors are stored in the status log. Errors are stored in the status log for 180
days, even if they are corrected.
Alarm field
The alarm field on the display will always flash when an error is pending.
Error output
The digital and relay output can be activated individually error by error (error level). The relay
output is default selected to error level. An output can also be selected to activate on a single
error number.
The alarm field, error output and error pending always operate together.
The analog output turns to a 1 mA level when in the 4 ... 20 mA mode.
Operator menu
Error pending and status log are as default enabled ( ) in the operator menu.