Additional Installation Instructions
B.4 Hi Precision mount
FSS200 clamp-on sensors
Installation Manual, 08/2017, A5E36255466-AC
131
7. Reinstall the Mylar template on the pipe and tape it together.
8. Align the marks on the template with the horizontal plane of the pipe. Align the strap
retainers, on center, with the marks on the Mylar template.
Figure B-47 Mylar template
9.
Use the Wizard setup procedure to program the transmitter for the sensors that were
selected.
10.Check index spacing on the flowmeter and adjust strap retainers for the correct spacing
between the sensors.
B.4.4.3
Direct mount enclosure housings
Installation procedure
1. Install the first enclosure housing and align it with the center line of the Mylar template.
2. Fully tighten and torque all stand off strap retainer nuts to approximately 2.25 - 2.8
newton/meters (20-25 in/lbs).
Lock stand offs in place with second nut.
Note
Maintain the alignment of the stand offs throughout the remaining installation procedures.
This is the reference point used when mounting the sensor housing and aligning the
remaining stand offs and strap retainers
.
3. Loosely install the second enclosure housing. Position it 180° from the Mylar template
center line.
4. Fully tighten and torque all stand off strap retainer nuts to approximately 2.25 -2.8
newton/meters (20-25 in/lbs).
Lock stand offs in place with second nut.