Connecting
5.5 Connecting sensors to transmitter
FSS200 clamp-on sensors
66 Installation Manual, 08/2017, A5E36255466-AC
5. Push the right-angle F-connector onto to the connector on the transmitter PC board.
6. Screw the gland and nut together to secure the cables.
Note
Sensor ports are labeled A and B.
Right-angle F-connector
F-connector
I/O cable gland
5.5.4
HI Precision mount FST030
Single enclosure Reflect mount wiring
Connect sensor cables to the transmitter as follows:
1. Fill connector end with Super Lube prior to connecting.
2. Apply Super Lube to the internal threads at the large end of the thread connector.
3. Observing the upstream and downstream orientation
, connect upstream sensor cable
to transmitter port (1A) and make connection snug.