Installing/mounting
4.6 Installation instructions
FSS200 clamp-on sensors
28 Installation Manual, 08/2017, A5E36255466-AC
Upstream sensor
Downstream sensor
Pipe
Sonic path
Flow direction
Figure 4-4 Direct mount (Pipe shown from above in 12 o'clock position)
Mounting supplies
The following items will be needed to mount the sensors (most materials are supplied):
Flat blade screwdriver
Mounting frames or mounting tracks
Tape, chalk and a ruler or measuring tape
Mounting straps
Spacer bar
Mounting guide (for Direct mount)
Ultrasonic coupling compound and/or coupling pads
Sensors (matched set)
Mounting strap kits
The available mounting strap kits are listed below. Each kit comes with up to two band sizes
to cover its designated pipe diameter range and a spacing guide for Direct mount. In most
cases, the kits come with the mounting hardware.
Mounting strap kits
Pipe diameter
SAE band sizes (Qty.)
7ME396000SM00
50.8 mm (2-inch) to 177.8 mm (7-inch)
#88 (2) #128 (2)
7ME396000SM10
50.8 mm (2-inch) to 330.2 mm (13-inch)
#88 (2) #152 (2)
7ME396000SM20 330.2 mm (13-inch) to 609.6 mm (24-
inch)
#188 (2) #280 (2)
7ME396000SM30 609.6 mm (24-inch) to 1219.2 mm (48-
inch)
#152 (4) #312 (4)