Commissioning
3.3 Setting the Parameters
FUP1010 IP67 Portable Quick Start
Operating Instructions, 01/2013, CQO:QSG003 Revision 05
19
4. Press <ENTER>. Enter dimensions manually if pre-programmed dimensions do not
match application.
Note
The DN sizes listed in the [Select Pipe Size] menu option list are referenced to DIN Table
2448. After selecting pipe size, check pipe OD and wall thickness for correct dimensions.
5. Press the <Left Arrow> and return to the main menu.
Select Liquid Class
1. Press the <Down Arrow> and scroll to [Application Data].
2. Press the <Right Arrow> to select [Liquid Class].
3. Press the <Right Arrow> again and scroll to desired liquid.
4. Press <ENTER> to save selection.
ī˜”
6HOHFW/LTXLG&ODVVIURP/LTXLG7DEOH
/LTXLG&ODVV :DWHUī˜ƒī˜•ī˜“&ī˜’ī˜™ī˜›)
8QL0DVV7DEOH 'LVDEOHG
7HPSī˜‘ī˜ƒ5DQJHī˜ƒī˜ƒī˜ƒī˜ƒī˜ƒī˜ƒī˜ƒī˜ƒī˜ƒī˜ƒī˜ƒī˜ƒī˜ƒī˜ƒī˜ƒī˜ƒī˜ƒī˜ī˜—ī˜“)WRī˜ƒī˜•ī˜˜ī˜“)
3LSH&RQILJ)XOO\'HYHORSHG
$QRPDO\'LDPV ī˜ƒī˜ƒī˜ƒī˜ƒī˜ƒī˜ƒī˜ƒī˜ƒī˜ƒī˜ƒī˜”ī˜“
$SSOLFDWLRQ'DWD
6LHPHQVī˜ƒī˜ƒī˜ƒī˜ƒī˜ƒī˜ƒī˜ƒī˜ƒī˜ƒī˜ƒī˜ƒī˜•ī˜ƒ&KDQQHO>ī˜”@$%&
ā‘  Select from list.
Select Pipe Configuration
1. Scroll down to [Pipe Config] and press the <Right Arrow>.
2. Select a configuration that approximates the conditions upstream of your Sensor
mounting location. (Refer to the definitions below.)
3. Press <ENTER> to save selection.