INSTALLATION 3
19
SITRANS FX300
www.siemens.com/flow06/2009 - SFIDK.PS.050.F2.02 A5E02100423
3.4.1 Measurement of liquids
Prohibited installation
Prohibited installationProhibited installation
Prohibited installation
Figure 3-2: Upper pipe bend
CAUTION!
Prohibited: Installing the device in an upper pipe bend 1, because there is a risk of gas bubbles
2 forming. Gas bubbles can lead to pressure surges and inaccurate measurement.
Figure 3-3: Downpipe and outlet
CAUTION!
Prohibited: Installing the device in a downstream pipe 3 or upstream pipe of a outlet 4. There is
the risk of partially filled pipes.
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 19 Monday, June 22, 2009 8:17 AM