SERVICE 7
63
SITRANS FX300
www.siemens.com/flow06/2009 - SFIDK.PS.050.F2.02 A5E02100423
Service
7.1 Replacing signal converter / LCD display
The signal converter has to be replaced by a converter of the same type. The following
parameters have to be observed for it:
The item number must match: 2.143670.xxx
The software version 2 must match.
The basic version has no software identification
The gas version is marked: "gas"
The steam version is marked: "steam"
The following steps are to be performed:
Switch off the power supply.
Unscrew the front cover.
Detach the display from the spacer pins.
Detach the display cable 1.
Detach the measuring sensor cable 3.
Loosen the three fixing screws 5 (Phillips).
Pull out signal converter.
Insert new signal converter.
Tighten the three screws.
Attach measuring sensor cable 3 (Caution: ensure that the measuring sensor cables are not
on top of the spacer pins 4 May be damaged).
Attach display cable 1.
Attach display in desired position, apply even pressure to the entire surface.
Screw on cover by hand.
Figure 7-1: Board, example
1 Connector for display
2 Version number, software identification
3 Connector for measuring sensor
4 Spacer pins
5 Fixing screw (Phillips)
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 63 Monday, June 22, 2009 8:17 AM