TECHNICAL DATA 8
69
SITRANS FX300
www.siemens.com/flow06/2009 - SFIDK.PS.050.F2.02 A5E02100423
Process connections
Flange version
Flange versionFlange version
Flange version
DIN EN 1092-1 DN15...300 in PN16100
ASME B16.5 ½...12" in 150600 lbs
JIS B 2220 DN15...300 in JIS 1020 K
For detailed information on combination flange/pressure rating,
refer to chapter "Dimensions and weights".
Sandwich version
Sandwich versionSandwich version
Sandwich version
DIN DN15...100 in PN100 (higher pressures on request)
ASME ½...4" in 600 lbs (higher pressures on request)
JIS DN15...100 in 1020 K (higher pressures on request)
Electrical connections
Power supply Non-Ex: 1436 VDC
Ex: 1430 VDC
Current output
Current outputCurrent output
Current output
Description of the used
abbreviations
U
ext
= external voltage; R
L
= load + resistance
Measuring range
4...20 mA (max. 20.8 mA) + HART
®
protocol
Load Minimum 0 Ω; maximum RL = ((Uext. - 14 VDC) / 22 mA)
Error signal Acc. to NAMUR NE43
Upper value: 21.0 mA
Lower value: 3.6 mA (not with HART
®
protocol)
Pulse output
Pulse outputPulse output
Pulse output
Pulse rate Max. 0.5 pulse/s (corresponds to 1800 pulses/hour)
Power supply Non-Ex: 24 VDC as NAMUR or open < 1 mA, maximum 36 V, closed
100 mA, U < 2 V
Ex: 24 VDC as NAMUR or open < 1 mA, maximum 30 V, closed 100 mA,
U < 2 V
HART
HARTHART
HART
®
HART
®
protocol via current output
Device revision 1
Physical layer FSK
Device category Transmitter, galvanically isolated
System requirements Load min. 250 Ω
Multidrop operation 4mA
Approvals and certifications
ATEX ATEX II 2G EEx d ia [ia] IIC T6
FM Class I, II, III
Div 1/2, groups A-G
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 69 Monday, June 22, 2009 8:17 AM