7ML19985ANO2 SITRANS LU10 – INSTRUCTION MANUAL Page 103
mmmmm
Specifications
Specifications
Power: • 100 / 115 / 200 / 230 Vac ± 15%, 50 / 60 Hz, 31 VA
Environmental: • location :
• altitude :
• ambient temperature :
• relative humidity :
• installation category :
• pollution degree :
• indoor / outdoor
• 2000 m max
• – 20 to 50 °C (– 5 to 122 °F)
• suitable for outdoor
(Type 4X / NEMA 4X / IP65 enclosure)
•II
•4
Scan Points: • 10 points per SITRANS LU10 max
• frequency independent
Range: • 0.3 m (1 ft) to 60 m (200 ft) max
Accuracy:
• 0.25% of program range
1
or 6 mm (0.24"), whichever is greater
1.
Program range is defined as the empty distance to the face of the transducer (P006) plus any range
extension (P801).
Resolution:
• 0.1% of program range
1
or 2 mm (0.08"), whichever is greater
Memory: • EEPROM (non-volatile) no back-up battery required
Programming: • via removable hand programmer and / or Dolphin interface package
Display: • Custom Graphics LCD
• 51 mm (2 in) x 127 mm (5 in) viewing area
Synchronization • up to 16 SITRANS LU10 units can be synchronized together
Temperature
Compensation:
• – 50 to 150 °C (– 50 to 302 °F)
• integral transducer sensor
• 1 TS-3 temperature sensor input (expandable to 10 inputs with optional
TIB-9 card).
• programmable fixed temperature
Temperature
Error:
•with compensation:
• fixed temperature:
• 0.09% of range
• 0.17% / °C deviation from programmed
temperature.
Outputs: • transducer drive: • 150 to 315 V peak (transducer model
dependent)
Communications: • proprietary bipolar current loop for Peripherals communication (see
Options)
• Dolphin compatible
• SmartLinx compatible
Enclosure:
• Type 4X / NEMA 4X / IP65
2
• 285 mm W x 209 mm H x 92 mm D (11.2" W x 8.2" H x 3.6" D)
• polycarbonate
2.
The use of approved watertight hubs/glands is required for Type 4X / NEMA 4X / IP65.
Weight: • 2.3 kg (5 lb)
Approvals: • CE *, CSA
NRTL/C
• Lloyd’s Register of Shipping, Categories ENV1, ENV2, ENV3, and ENV5
* EMC performance available upon request