Page 44 SITRANS LU10 – INSTRUCTION MANUAL 7ML19985AN02
mmmmm
Enhancement Parameters
Profile Records (P330 to P337)
The following parameters are for authorized Siemens Milltronics service personnel or
Instrumentation Technicians familiar with Siemens Milltronics echo processing
techniques.
Use these features to record and save a total of up to 10 Echo profiles, initiated manually
(P330), or automatically (P331 et al). See Scope displays (P810) for echo profile viewing
hardware / software requirements.
If 10 Echo Profiles are already saved, addresses 1 through 10 are filled, the oldest
automatically initiated record is overwritten. Manually initiated records are not
automatically overwritten. All records are automatically deleted in the event of a power
interruption.
When a record is displayed, results are based on current programming (which may have
been altered since the record was saved). This permits the effect on the echo profile to be
observed when changing an echo parameter.
P302 (V) TEMPERATURE, SENSOR MAX
Use this feature if vessel temperature is monitored by a TS-3 Temp Sensor to
view the highest temperature recorded (in °C).
When accessed, the TS-3 symbol is displayed in the Point Type field and the
TS-3 Number (corresponding to the SITRANS LU10 terminals) is displayed in the
Point Number field.
values: -50 to 150