7ML19985AN02 SITRANS LU10 – INSTRUCTION MANUAL Page 27
mmmmm
Application Parameters
Application Parameters
Application Parameters are used to alter the SITRANS LU10 display, failsafe, relay, and /
or mA output operation.
Volume Parameters (P050 to P055)
If Readings proportional to volume are desired, adjust the following parameters.
If Volume Conversion is not required, proceed to Reading Parameters.
If Volume Conversion is not required, proceed to Reading Parameters. If Volume Conversion is not required, proceed to Reading Parameters.
If Volume Conversion is not required, proceed to Reading Parameters.
P050 TANK SHAPE
Enter the Tank Shape option that matches the vessel monitored.
If the Tank Shape option selected requires additional vessel dimension entry, the
associated parameters
associated parametersassociated parameters
associated parameters (as indicated below) may be scroll accessed.
When Operation is "level" (P001 = 1), material volume is calculated. Alternatively,
when Operation is "space" (P001 = 2), remaining vessel capacity is calculated.
In the RUN mode, Readings are displayed in percent of (and mA outputs are
proportional to) maximum volume. To convert Readings to volumetric units, see
Max Volume (P051).
When monitoring solids, volume calculation accuracy may vary.
values: 0 = volume calculation not required (preset)
1 =
flat level bottom
or
A
4 =
half sphere bottom (P052)
8 =
sphere
A
or
2 =
cone / pyramid bottom (P052)
5 =
A
or
flat slope bottom (P052)
9 =
Universal Linear (P054 / P055)
3 =
A
parabola bottom (P052)
6 =
flat ends
7 =
A
L
parabola ends (P052 / P052)
10 =
Universal Curved (P054 / P055)