Page 28 SITRANS LU10 – INSTRUCTION MANUAL 7ML19985AN02
mmmmm
Application Parameters
* Steps to enter a Level Breakpoint or Breakpoint Volume:
1.. Press to display the index symbol, .
2. Scroll ( or ) or direct access the desired Breakpoint.
3. Key in the desired level or volume value.
4. Press .
Refer to
Refer to Refer to
Refer to
Technical Reference
Technical Reference Technical Reference
Technical Reference
Volume Calculation for breakpoint value selection.
Volume Calculation for breakpoint value selection. Volume Calculation for breakpoint value selection.
Volume Calculation for breakpoint value selection.
P051 MAX VOLUME
Use this feature to display the Reading in volumetric units rather than percent.
Enter the vessel volume between Empty (P006) and Span (P007).
e.g. 1
If the volume = 3650 m
3
, enter 3650.
e.g. 2 If the volume = 267,500 U.S. gallons, enter 267.5 (1000’s of gallons).
values: 0.000 to 9999
P052 TANK DIMENSION A
Enter the height of the tank bottom if P050 = 2,3,4, or 5, or the length of one end
section of the tank if P050 = 7, in Units (P005).
values: 0.000 to 9999
P053 TANK DIMENSION L
Enter the length of the tank (excluding both end sections) if P050 = 7, in Units
(P005).
values: 0.000 to 9999
P054 LEVEL BREAKPOINTS ( Universal Volume Calculation )
Enter level breakpoints* (where volume is known) if P050 = 9 or 10
values: 0.000 to 9999
P055 BREAKPOINT VOLUMES ( Universal Volume Calculation )
Enter the volume* corresponding to each Level Breakpoint entered.
values: 0.000 to 9999