Page 72 SITRANS LU10 – INSTRUCTION MANUAL 7ML19985AN02
mmmmm
Enhancement Parameters
Test Parameters (P900 to P913)
Test Parameters are intended for use by Siemens Milltronics Service personnel.
P900 (V) SOFTWARE REV. #
Access this parameter to determine the EPROM Rev. # without removing the
enclosure lid.
values: 00.00 to 99.99
P901 (V) MEMORY
Press to activate the SITRANS LU10 memory test.
When the test is successful, "PASS" is displayed. Otherwise, one of the following
values is displayed, indicating the type of memory failure.
values: PASS = memory test successful
F1 = RAM
F2 = NOVRAM
F3 = EEPROM
F4 = EPROM
P902 (V) WATCHDOG
Press to reset the microprocessor.
On successful completion (in about 10 seconds) the SITRANS LU10 enters the
RUN mode.
P903 (V) DISPLAY
Press to activate the display test.
All LCD segments and symbols are temporarily displayed.
P904 (G) KEYPAD
Press each keypad key in the following sequence:
As each key is pressed, the associated keypad number is displayed. On
successful test completion "PASS" is displayed. "FAIL" is displayed if a key is
pressed out of sequence or the programmer keypad malfunctions.
1 2 3 4
mA
5 6 7 8
9 0
P
P