Page EN-4 XPS/XCT Series Transducers – OPERATION MANUAL A5E32282889
mmmmm
English
The certification of this equipment relies on the following materials used in their
construction:
For manual override, use the disconnect switch provided in the building installation of the
associated controller.
XPS 30/40 Series – Certificate SIRA 01ATEX5153X
This equipment may be used in hazardous dust zones with all conductive and non-
conductive dusts. The XPS-30 and XPS-40 type series transducers have a maximum
surface temperature of 135 °C (275 °F) (Temperature Class T4). These units are certified
for use in ambient range of -40 to 95 °C (-40 to 203 °F). The transducers should not be
used outside this temperature range. The XPS-30 and XPS-40 ultrasonic transducers
must be installed so the face of the transducer is not substantially subjected to light.
Installation shall be carried out in accordance with the applicable code of practice, and
by suitably trained personnel. Repair of the equipment shall be carried out in accordance
with the applicable code of practice and installation instructions.
These devices should only be supplied from a circuit containing a suitably rated fuse that
has a breaking capacity of 4000A. This fuse is included in Siemens Milltronics controllers.
The certification of this equipment relies on the following materials used in their
construction:
For manual override, use the disconnect switch provided in the building installation of the
associated controller.
Product Marking
XPS Series XCT Series
Enclosure Kynar®
1
710
1.
Kynar®is a registered trademark of ELF Atochem.
Durapot®is a registered trademark of Cotronics Corporation.
Kynar®710
Encapsulant Stycast LA-9823-76 Durapot® 861-F3 & 864
Enclosure: Kynar®710
Encapsulant: Stycast LA-9823-76
Note: Kynar®polyvinylidene flouride is resistant to attack from most chemicals under the
described operating conditions. However, for exposure to specific environments, check
with chemical compatibility charts prior to installation.
WARNING: This product is designated as a Pressure Accessory per Directive 97/
23/EC and is
not
intended for use as a safety device.