4.1.5. Factory Default Settings
Communication schedule: Each hour
Measurement standard: 1A (SS 460 48 66 Spräng, 250mm/s)
Threshold: 2.0 mm/s (ON)
Interval time: 2 minutes
Record time: 4 seconds
GPS timesync: Disabled
4.2. Mounting
1. Identify an even surface where the C22 can be mounted.
2. Drill for the fastener (e.g. a concrete anchor).
3. Insert fastener.
4. Screw a nut to the head of the bolt, keep it loose but close to head
of the screw.
5. Attach a washer to the bolt/nut.
6. Insert the bolt through the center hole of the C22 and screw the
bolt into the anchor (C22 hanging loose).
7. Hold the head of the bolt still with one tool and use an appropriate
tool to tighten the nut to secure the C22 to the wall.
If the wall or ground is not completely flat, you can adjust one or two of
the mounting screws to ensure all three have full contact with the wall.
Make sure that the C22 is firmly attached and aligned (plumb and level).
For correct measuring results, the mounting angle of the instrument must
not exceed 5º relative to the horizontal plane. Refer to Sigicom product
catalog for more information on mounting solutions.
Note! The instrument must not be submerged in water.