Laboratory centrifuge Sigma 1-14
3 Safety
18 / 67
Pos: 43 /100 Sig ma/100 BA Zentrif ugen Sigma (S tandardmod ule)/030 Sicherheit /030-0060 Sicherheitshin weise------------------------------------------------------------- @ 26\mod_140531902 1480_68.docx @ 192031 @ 2 @ 1
3.6 Safety instructions
Pos: 44 /100 Sig ma/100 BA Zentrif ugen Sigma (S tandardmod ule)/030 Sicher heit/030-0060-0 010 Elektrisch e Sicherheit @ 26\mod_14 05319022891_68. docx @ 192 045 @ 3 @ 1
3.6.1 Electrical safety
To reduce the risk of electrical shock, the centrifuge uses a three-wire
electrical cord and plug to connect the equipment to earth-ground. To
preserve this safety feature:
DANGER
Ensure that the wall socket is properly wired and grounded.
Check that the mains voltage agrees with the nominal voltage listed on
the name plate.
Do not place vessels containing liquid on the centrifuge lid or within the
safety distance of 30 cm around the centrifuge. Spilled liquids may get
into the centrifuge and damage electrical or mechanical components.
Work on the power supply system must only be performed by certified
electricians.
Inspect the electrical equipment of the unit regularly. Defects such as
loose or burnt cables must be eliminated immediately.
Pos: 45 /010 Uni versalmodule/ Leerzeile @ 0\mod_120 2116244500_ 0.docx @ 114 @ @ 1
Pos: 46 /100 Sig ma/100 BA Zentrif ugen Sigm a (Standardmod ule)/030 Sicher heit/030-0060-0 020 Mechanisc he Sicherh eit (außer 1-1 6/1-16K) @ 26\ mod_140531902 4314_68.docx @ 192059 @ 3 @ 1
3.6.2 Mechanical safety
In order to ensure the safe operation of the centrifuge, observe the
following:
WARNING
Do not open the lid when the rotor is in motion!
Do not reach into the rotor chamber when the rotor is in motion!
Do not use the centrifuge if it was installed incorrectly.
Do not use the centrifuge without panels.
Do not use the centrifuge if the rotors and inserts show signs of
corrosion or other defects.
Only use the centrifuge with rotors and accessories that have been
approved by the manufacturer. In case of doubt, contact the
manufacturer (see chapter 7.3 - "Service contact").
Do not hold your fingers between the lid and the housing when closing
the lid. Risk of crushing!
Defective lid relieving devices could cause the centrifuge lid to fall
(contact the service department, if necessary). Risk of crushing!
Do not hit or move the centrifuge during its operation.
Do not lean against or rest on the centrifuge during its operation.
Do not spin any substances that could damage the material of the
rotors and buckets of the centrifuge in any way. Highly corrosive
substances, for example, damage the material and affect the
mechanical strength of the rotors and buckets.
Stop the centrifuge immediately in the event of a malfunction. Eliminate
the malfunction (see chapter 7 - "Malfunctions and error correction") or
inform the service department of the manufacturer (see chapter 7.3 -
"Service contact").
Ensure that all repairs are performed only by authorised and
specialised personnel.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sigma 1-14 and is the answer not in the manual?

Sigma 1-14 Specifications

General IconGeneral
BrandSigma
Model1-14
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals