Laboratory centrifuge Sigma 1-14
3 Safety
Version 08/2012, Rev. 3.14 of 11/08/2017 sb
21 / 67
Translation of the original operating manual, part no. 0700002
Pos: 56 /100 Sig ma/100 BA Zentrif ugen Sigma (S tandardmod ule)/030 Sicherheit /030-0060-0 070 Sicherheit von Rotoren und Zubehör @ 44\mod_147 0119200568_ 68.docx @ 437008 @ 3 @ 1
3.6.7 Safety of rotors and accessories
Pos: 57 /100 Sig ma/100 BA Zentrif ugen Sigma (S tandardmod ule)/030 Sicherheit /030-0060-0 070-0010 L ebensdauer @ 2 6\mod_14053 19032612_68.do cx @ 192143 @ 3 @ 1
3.6.7.1 Service life
The rotors and accessories have a limited service life.
WARNING
Perform regular checks (at least once per month) for safety reasons!
Pay special attention to changes, such as corrosion, cracks, material
abrasion, etc.
After 10 years, they must be inspected by the manufacturer.
After 50,000 cycles, the rotor must be scrapped for reasons of safety.
If other data concerning the service life are engraved on the rotor or
bucket, these data shall apply accordingly. For example, a bucket with
the engraving “max. cycles = 10,000” has a service life of 10,000
cycles, and a rotor with the engraving “Exp. date 02/20” must be
scrapped in February 2020 at the latest (see figure).
Fig. 2: Different service life engraving on the bucket/rotor
NOTE
Refer to the table of rotors and accessories with a different service
life (see chapter 11.3 - "Table of the service life of rotors and
accessories")!