Laboratory centrifuge Sigma 1-14
8 Maintenance and service
Version 08/2012, Rev. 3.14 of 11/08/2017 sb
41 / 67
Translation of the original operating manual, part no. 0700002
Pos: 155 /100 Sig ma/100 BA Ze ntrifugen Sigma (St andardmo dule)/080 Wartung und Instandh altung/080-0 010-0020 Zu behör (außer 8K) @ 28\mod_14 05320618247_ 68.docx @ 1 94845 @ 3 @ 1
8.1.2 Accessories
CAUTION
For the care of the accessories, special safety measures must be
considered as these are measures that will ensure operational safety at
the same time!
Immediately rinse off the rotor, buckets, or accessories under running
water if they have come into contact with any liquids that may cause
corrosion. Use a brush for test tubes in order to clean the bores of angle
rotors. Turn the rotor upside down and allow it to dry completely.
Clean the accessories outside the centrifuge once a week or preferably
after each use. Adapters should be removed, cleaned and dried.
CAUTION
Do not clean the accessories in a dishwasher!
Cleaning in a dishwasher removes the anodised coating; the result is
cracking in areas that are subject to stress.
If the rotors or accessories have been contaminated with toxic,
radioactive, or pathogenic substances, clean them immediately with a
suitable decontamination agent (depending on the type of
contamination). Take suitable precautions for your own safety if there is
a risk of toxic, radioactive, or pathogenic contamination.
Dry the accessories with a soft cloth or in a drying chamber at
approx. 50°C.
Pos: 156 /010 Uni versalmodule/ Leerzeile @ 0\mod_12 02116244500 _0.docx @ 114 @ @ 1
Pos: 157 /100 Sig ma/100 BA Ze ntrifugen Sigma (St andardmo dule)/080 War tung und Insta ndhaltung/080-0 010-0020-001 0 Kunststoffz ubehör @ 28\ mod_1405320619 602_68.docx @ 194859 @ 4 @ 1
8.1.2.1 Plastic accessories
The chemical resistance of plastic decreases with rising temperatures (see
chapter 11.4 - "Resistance data").
If solvents, acids, or alkaline solutions have been used, clean the plastic
accessories thoroughly.
WARNING
Plastic accessories must not be greased!
Pos: 158 /010 Uni versalmodule/ Seitenwechsel @ 0\mod_12 02116244312_0 .docx @ 105 @ @ 1

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sigma 1-14 and is the answer not in the manual?

Sigma 1-14 Specifications

General IconGeneral
BrandSigma
Model1-14
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals