Laboratory centrifuge Sigma 1-14
Version 08/2012, Rev. 3.14 of 11/08/2017 • sb
Translation of the original operating manual, part no. 0700002
Pos: 52 /100 Sig ma/100 BA Zentrif ugen Sigma (S tandardmod ule)/030 Sicherheit /030-0060-0 050 Sicherheits hinweise zur Z entrifugation @ 26\mod_140 5319029632_6 8.docx @ 19 2115 @ 3 @ 1
3.6.5 Safety instructions for centrifugation
For safe operation, observe the following before starting the centrifuge:
• Ensure that the centrifuge was set up properly (see chapter 5 - "Set-up
and connection").
• Maintain a safety distance of at least 30 cm (12 inches) around the
centrifuge.
• Do not store any dangerous goods in the centrifuge area.
• Do not stay in the safety area longer than what is absolutely necessary
for the operation of the centrifuge.
• Only use the centrifuge with rotors and accessories that have been
approved by the manufacturer. We explicitly warn against the use of
equipment of poor quality. Breaking glass or bursting vessels can
cause dangerous imbalances at high speeds
• Ensure that rotor and buckets are correctly fitted (see chapter 6.2.2.1 -
"Installation of the rotor").
• Observe the instructions on the installation of accessories (see chapter
6.2.2.3 - "Installation of accessories").
• The rotor must be loaded axial symmetrically at equal weights.
• If liquids with a density > 1.2 g/cm
3
are used, reduce the speed (see
chapter 2.2.2.2 - "Density").
• Do not use the centrifuge if the rotor is loaded asymmetrically.
• Do not use the centrifuge with tubes that are excessively long.
Pos: 53 /010 Uni versalmodule/ Leerzeile @ 0\mod_120 2116244500_ 0.docx @ 114 @ @ 1
Pos: 54 /100 Sig ma/100 BA Zentrif ugen Sigma (S tandardmod ule)/030 Sic herheit/030-0060-0 060 Bestän digkeit von Kuns tstoffen @ 26\ mod_140531 9031161_68.doc x @ 192129 @ 3 @ 1
3.6.6 Resistance of plastics
Chemical influences have a strong effect on the polymeric chains of
plastics, and, therefore, on their physical properties. Plastic parts can be
damaged if solvents, acids, or alkaline solutions are used.
• Refer to the resistance data (see chapter 11.4 - "Resistance data")!