6.3.7 Runtime
The preselected total runtime is displayed in the lower line of the display.
During centrifugation, the remaining runtime is displayed. The runtime of
the centrifuge can be set at one-second-intervals up to 99 minutes and 59
seconds.
• To select a runtime, press the select key repeatedly until the time unit
flashes on the display.
• Select the desired runtime by pressing the arrow keys.
• Confirm the selection by pressing the start/stop key.
The data will also be accepted and saved after approx. 20 seconds if no
other key is pressed within this time period.
Pos: 124 /010 Uni versalmodule/ Leerzeile @ 0\mod_12 02116244500 _0.docx @ 114 @ @ 1
Pos: 125 /100 Sig ma/110 BA Ze ntrifugen Sigma (Proj ekte)/1-14_1- 14K/061 Be trieb/Spincontr ol Basic/061-0 030-0070-001 0 Ändern der L aufzeit während des Laufs (nur 1- 14) @ 28\ mod_1405320829 604_68.doc x @ 195601 @ 4 @ 1
6.3.7.1 Changing the runtime during centrifugation
The preselected runtime can be changed during centrifugation.
• Press the select key repeatedly until the time unit flashes on the display.
• Change the desired runtime by pressing the arrow keys. The
parameters will be accepted immediately.
If the centrifugation time is changed during the run, the centrifuge will run
for the entire new time and will disregard the previous runtime that has
already elapsed.
Pos: 126 /010 Uni versalmodule/ Leerz eile @ 0\mod_12 02116244500 _0.docx @ 114 @ @ 1
Pos: 127 /100 Sig ma/100 BA Ze ntrifugen Sigma (St andardmo dule)/061 Betrieb_ Spincontrol Basic /Spincontr ol Basic /061-0 030-0070-002 0 Kurzzeitbetri eb (nur 1-14, 1-1 4K) @ 28\m od_14053210 80308_68.doc x @ 196091 @ 4 @ 1
6.3.7.2 Short run (quick run)
During the short run, the centrifuge accelerates at maximum power until the
maximum speed is reached.
• Keep the start/stop key pressed during the short run.
The message "short run" and the duration of the short run is displayed.
When the start/stop key is released, the centrifuge decelerates at maximum
power to a standstill.
After the short run, the lid unlocks automatically and the program that was
set beforehand is displayed again.