Laboratory centrifuge Sigma 1-14
8 Maintenance and service
Version 08/2012, Rev. 3.14 of 11/08/2017 • sb
Translation of the original operating manual, part no. 0700002
Pos: 167 /100 Sig ma/100 BA Ze ntrifugen Sig ma (Standardmo dule)/080 Wartung und Instandh altung/080-0 020-0010 A utoklavieren @ 28\mod_140532 0627440_68. docx @ 19494 3 @ 3 @ 1
8.2.1 Autoclaving
The service life of the accessories essentially depends on the frequency of
autoclaving and use.
• Replace the accessories immediately when the parts show changes in
colour or structure or in the occurrence of leaks etc.
• During autoclaving, the caps of the tubes must not be screwed on in
order to avoid the deformation of the tubes.
It cannot be excluded that plastic parts, e.g. lids or carriers, may deform
during autoclaving.
Polyallomer / polycarbonate rectangular carriers
Polyallomer / polycarbonate round carriers
Polycarbonate / polyallomer lids for angle rotors
Polyphenylsulfone (PPSU) caps for buckets
Polypropylene balance adapter for blood-bag systems
Polypropylene copolymer tubes
Polypropylene rectangular carriers
Polypropylene round carriers
Polysulfone caps for buckets
Polysulfone lids for angle rotors
Stainless-steel balance weight for blood-bag systems