7 Malfunctions and error correction
Pos: 140 /100 Sig ma/100 BA Ze ntrifugen Sigma (St andardmo dule)/070 Störu ngen und Fehlersu che/070-00 10 Allgemein e Störungen - B asic (außer 2-7) @ 28\mod _1405320577437 _68.docx @ 19 4677 @ 2 @ 1
7.1 General malfunctions
Malfunctions are indicated by error messages with a number in the
speed/RCF display.
In the event of a fatal error (e.g. a defective lid lock), a certain safety time
will be counted down on the display. During this time, “ERR” and “SAFE”
flash alternately on the display. When the time is up, "OFF" will be
displayed.
Do not switch the centrifuge off until “OFF” is displayed! This is necessary
in order to ensure that the rotor is at a complete standstill.
• Eliminate the source of the problem (see tables below).
• Acknowledge the error messages with the lid key.
No indication on the display
No power in the mains supply
Check fuse in the mains supply
Power cord is not plugged in
Plug in power cord correctly
Switch mains power switch on
Lid is not closed correctly
Centrifuge cannot be started: The set
speed value is displayed in an
unchanged manner
Netz aus/ein. Falls sich der Fehler
wiederholt, Service verständigen
The lid lock is not closed correctly
Open and close lid. If the error
occurs again, contact service
Centrifuge decelerates during
operation and displays an error from 1
to 18 after powering on
Power off/on. If the error occurs
again, contact service
Centrifuge decelerates during
operation and displays error 19 after
powering on
Quit by pressing the lid key
Temperature value cannot be reached
(only for refrigerated centrifuges)