Laboratory centrifuge Sigma 1-14
8 Maintenance and service
42 / 67
Pos: 159 /100 Sig ma/100 BA Ze ntrifugen Sigma (St andardmo dule)/080 Wartung und Instandh altung/080-0 010-0031 Rotor en (nur Basic) @ 28\mod_1 40532062233 6_68.docx @ 194 887 @ 3 @ 1
8.1.3 Rotors
The rotors are produced with the highest precision, in order to withstand the
permanent high stress from high gravitational fields.
Chemical reactions as well as stress-corrosion (combination of oscillating
pressure and chemical reaction) can affect or destroy the metals. Barely
detectable cracks on the surface can expand and weaken the material
without any visible signs.
Check the material regularly (at least once a month) for
cracks
visible damage of the surface
pressure marks
signs of corrosion
other changes.
Check the bores of the rotors and multiple carriers.
Replace any damaged components immediately for your own safety.
Pos: 160 /010 Uni versalmodule/ Leerzeile @ 0\mod_12 02116244500 _0.docx @ 114 @ @ 1
Pos: 161 /100 Sig ma/110 BA Ze ntrifugen Sigma ( Projekte)/1-1 4_1-14K/080 Wartung und Ins tandhaltung/0 80-0010-0040 Mi krohämatokrittell er 1-14_1-1 4K @ 28\mod_ 14053208628 86_68.docx @ 195685 @ 3 @ 1
8.1.4 Microhaematocrit rotor
Remove the microhaematrocrit rotor for cleaning.
Wipe the centrifuge chamber clean.
Replace the rubber ring in the event of wear or glass breakage (part no.
16002 for rotor 11026).
Pos: 162 /010 Uni versalmodule/ Seitenwechsel @ 0\mod_12 02116244312_0 .docx @ 105 @ @ 1

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sigma 1-14 and is the answer not in the manual?

Sigma 1-14 Specifications

General IconGeneral
BrandSigma
Model1-14
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals