Laboratory Centrifuge Sigma 6-16S
6 Using the centrifuge
Version 07/2012, Rev. 1.22 of 08/05/2020 sb
51 / 91
Translation of the original operating manual, part no. 0702002
Restrictions
Acceleration and deceleration curves can include intervals with a
positive slope as well as intervals with a negative slope and also
intervals with 0 slope.
The slope of the curve intervals can be 1 min
-1
/sec (rpm/sec) minimum
and 1000 min
-1
/sec (rpm/sec) maximum.
Quadratic curve intervals are only possible between 0 and 1000 min
-1
(rpm) maximum. If a final speed > 1000 min
-1
(rpm) is selected, this
interval will automatically become linear above 1000 min
-1
(rpm).
The possible runtime results from the maximum possible speed
(depending on the rotor) and from the slope limitation.
Example 1: Start speed 0 min
-1
(rpm), final speed 100 min
-1
(rpm),
runtime 1 hour not possible, since the necessary slope < 0.03 min
-1
(rpm), which means that it is beyond the defined range.
Example 2: Start speed 0 min
-1
(rpm), final speed 15000 min
-1
(rpm),
runtime 10 sec. not possible, since the necessary slope is 1500 min
-1
(rpm), which means that it is beyond the defined range.
6.3.2.10 Help menu
The help function is symbolised by the " " symbol on the menu bar. It
provides a short description of the control elements of the selected option.
Activating and deactivating the help function
Select the question mark on the menu bar and press the function knob.
Quit the help function by selecting the question mark and by pressing
the function knob again.
Parameters can still be changed when the help function is activated.
Fig. 24: Menu "Help"

Table of Contents