Laboratory Centrifuge Sigma 6-16S
6 Using the centrifuge
Version 07/2012, Rev. 1.22 of 08/05/2020 sb
47 / 91
Translation of the original operating manual, part no. 0702002
6.3.2.8 Setup menu
The "Setup" menu is symbolised by the " " symbol on the menu bar. It is
used to perform basic settings concerning the control system of the
centrifuge. It enables the optimum adaptation of the centrifuge to its specific
area of application.
Fig. 22: Menu "Setup"
Function
Open lid after run
The automatic lid opening function ensures that the lid opens when the
rotor has stopped.
Program rotation
See chapter 6.3.3.5 - "Automatic program rotation"
Runtime as of set speed
If this function is active, the runtime will not be measured until the set speed
is reached. In the "Standard" menu, the symbol "!" will be displayed in the
runtime field.
NOTE
In the program mode, this function can be saved separately for every
program. In this case, the symbol "!" will be displayed under the runtime
of the detailed program description.
Start delay
If the start delay function is active, the centrifuge will not start until the
preset time has elapsed. The symbol "" will be displayed in the runtime
field.
Display deceleration time
In the activated mode, the deceleration time will be displayed instead of the
runtime during and after the deceleration process. Below the time display,
the symbol " " will be displayed. During a deceleration process, the
symbol flashes. Once the deceleration is complete, it is displayed in a
permanent manner.