Laboratory Centrifuge Sigma 6-16S
7 Malfunctions and error correction
Version 07/2012, Rev. 1.22 of 08/05/2020 sb
Translation of the original operating manual, part no. 0702002
7.1.1 Emergency lid release
In the event of a power failure, it is possible to manually open the
centrifuge lid.
Switch off the mains power switch and disconnect the power cord from
the socket.
Remove the plugs (see figure, item 1) from the opening on the left and
on the right side of the control panel, e.g. with a screwdriver.
Fig. 31: Position of the openings for the emergency lid release
Insert the supplied hexagon socket key horizontally into the hole. The
key will be guided through a funnel-shaped tube to the shaft of the lid
lock motor.
Fig. 32: The emergency lid release key must be inserted horizontally
Unlock the motorised lid locks as follows:
Turn the left lid lock anti-clockwise.
Turn the right lid lock clockwise.
Then, reinsert the plugs.
WARNING
Do not unlock or open the lid unless the rotor is at a standstill.
If the lid is opened via the emergency lid release system during a centrifuge
run, the centrifuge will be switched off immediately and decelerate in an
unbraked manner.