Operating manual - English UNOCODE PRO
Copyright Silca 2017 59
Y axis photocell
:
manually turn the Y axis disk
(fig. 36)
up
until the display’s description changes from OFF to ON
.
use the provided allen key to tighten the (G3) grub screw.
X axis photocell
:
manually turn the
X
axis disk
(fig. 34)
up
until the display’s description changes from OFF to
ON.
use the provided allen key to tighten the (G1) grub screw (fig. 34).
9) lower the protective shield
.
10)press
START.
- the machine goes on to a new position and is ready for regulating the sensors.
Fig. 36
Regulating the Y axis sensor
:
raise the protective shield.
use the provided allen key to loosen the (C1) grub screw; manually rotate the (C2) rod in both directions up until
the changeover point from OFF to ON is found.
tighten the (C1) grub screw to secure the rod.
Regulating the X axis sensor
:
carefully tip the machine over.
remove the bottom panel by unscrewing all 8 securing screws.
loosen the (C3) screw that secures the sensor support plate. Manually move the sensor support plate up until the
display’s description goes from OFF to ON.
tighten the (C3) screw to secure the plate back into place.
re-position the machine back on its workbench and lower the protective shield.
press START.
G3 G4
Y axis photocell
Y axis sensor