5Installation and startup Dentsply Sirona
5.4Commissioning Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
3. Add approx. 75 ml of DENTATEC to the tank.
4. Fill the tank with water until the filter insert is completely immersed
(up to the bottom edge of the cover thread, approx. 3 liters).
5. Wait for a short time until the filter insert is completely soaked; then
add an appropriate amount of water.
6. Close the water tank by tightening the tank cap clockwise by hand.
Do not use the tank cap opener for this.
7. Push the water tank back into the housing.
8. Switch the unit on (see Switching the unit ON and OFF [ → 32]).
9. Switch the pump on (press the
"Pump"
button) to fill the water circuit.
10. Fill the water tank up again until the filter insert is completely
immersed (up to the bottom edge of the cap thread).
5.4.9 Switching the unit ON and OFF
Switch on the unit
✔ The milling and grinding unit is connected to the power supply.
1. The main switch on the rear side of the unit is set to position I (ON).
2. Press the ON/OFF button on the front panel.
The unit switches on and the display lights up.
NOTICE
Damage to surfaces!
In the undiluted state, the coolant additive DENTATEC disintegrates
plastic surfaces and can cause discoloration.
➢ Do not place DENTATEC on the unit.
➢ Do not spill DENTATEC.
NOTICE
Do not put the unit into operation at low temperatures!
If you move the unit to the operating site from a cold environment,
condensation may form and result in a short circuit.
Within the machine, grease depots are included for lubricating
components that can cause error messages at low temperatures.
✔ Install the unit at room temperature.
➢ Wait until the unit has reached room temperature and is absolutely
dry (for at least one hour)
The unit is dry and can be put into operation.
NOTICE
Do not adjust the line voltage
The unit automatically adjusts to the line voltage.