63 98 650 D 3439
D 3439.201.07.10.23 10.2017
63
Dentsply Sirona 8Technical description
Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
8.2.2 Technical data
8.2.3 Controller board
● 3x 2-axis stepping motor controller with microstepping
● 2 DC motor controllers with integrated speed and current control and
force monitoring
● Ethernet, RJ45 interface 10 Mbit/sec
Type designation Milling and grinding unit CEREC
MC / MC X
Rated line voltage 100 V - 230 V AC
Rated power frequency 50/60 Hz
Nominal current 1.5 - 3.5 A
Nominal power output 320VA
Permissible line voltage fluctuations ±10% of nominal voltage
Type of protection against electric
shock
Class 1 equipment
Degree of protection against ingress
of water
Ordinary device (without
protection against ingress of
water)
Overvoltage category II
Ambient conditions For indoor use
Pollution degree 2
Air pressure: 700 hPa – 1060
hPa
Temperature range 5°C to 40°C
Humidity range 80% rel. up to 31 °C
decreasing to
50% rel. up to 40 °C
Operating mode Continuous operation
Dimensions (WxHxD) in mm 700 x 425 x 420
Approx. weight 43 kg