63 98 650 D 3439
52 D 3439.201.07.10.23 10.2017
7Service Dentsply Sirona
7.3Changing the water Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
To change the water, proceed as follows:
The unit is switched on.
No machining process is running.
1. Pull out the water tank at the front of the unit.
2. Open the drain opening.
3. Drain the water tank.
4. Turn the tank cap counter-clockwise and take it off. If you find the tank
cap hard to open by hand, use the tank cap opener (see "Opening the
tank cap").
5. Remove the filter insert from the tank and clean the filter thoroughly
under running water.
6. Rinse the water tank.
7. Insert the cleaned filter with handle into the tank and press it firmly
onto the base in the floor of the tank.
8. Close the drain opening.
9. Add approx. 75 ml of DENTATEC to the tank.
10. Fill the tank with water until the filter insert is completely immersed
(up to the bottom edge of the cover thread, approx. 3 liters).
11. Wait for a short time until the filter insert is completely soaked; then
add an appropriate amount of water.
12. Close the water tank by tightening the tank cap clockwise by hand.
Do not use the tank cap opener for this.
13. Push the water tank back into the housing.
NOTICE
Foaming not permissible!
If any cleaning agents are used, this will create foam, which is not
permitted.
Do not use any cleaning agents.