What to do if the Sirona Medical Equipment tip is worn, loose, or incorrect?
- Ssandra11Aug 30, 2025
If you're experiencing issues with the tip of your Sirona SIROPERIO 4 PS or SIROPERIO PE 8 equipment, consider the following: * For a worn tip, check its length with a tip gauge and replace it if necessary. * If the tip is loose, tighten it using the torque wrench. * If there is no tip, screw one on and tighten it with the torque wrench. * If you are using a tip other than the SIROPERIO 4 PS or SIROPERIO PE 8 tip, deactivate the detection function. * For a broken tip, screw on a new one and tighten it using the torque wrench. * Ensure the torque wrench is removed after tightening. * Verify the handpiece supply cable is correctly connected.