SITA Messtechnik GmbH Operating Instructions SITA ConSpector
27
Furthermore, it is recommended to have the SITA ConSpector and the optionally associated work-
ing standards serviced and adjusted by the manufacturer every 1-2 years as part of the monitoring
of the measuring equipment.
Table 1: Menu Checking
Menu Checking
Descripton
Last Check Date when the last successful device test was performed.
Perform Check
Performing the device check as a blank measurement and on the (optional) Calibra-
tion standards.
C0/C1
Setting whether a SITA calibration standard C1 is available (seria
l number and set
value). If no SITA calibraition standard C1 is available (optional), only an empty
measurement is performed during the test.
Warnung:
Setting when a new check should be requested. You can choose between the number
of days or the warning “after failure only” (lowest setting). Warning “after failure only”
means that you will be asked for a new test after an unsuccessful test.
Check limits
Limits for successfully passing a blank measurement test and limits for testing with the
C0 and C1 working standards.
Performing a test in the beaker (blank measurement)
Set up the measurement setup for the device test. The test is performed in the stainless steel beaker
with the lid closed. To perform a device test, select the "Perform check" menu item in the Test menu.
Then follow the on-screen instructions.
By pressing the software key
Continue, the SITA ConSpec-
tor starts with the empty
measurement in the beaker.
Blank measurement in beaker
Set value 0 RFU
Standard preset per-
missible tolerance
range
-20 RFU bis 50 RFU
These are basic settings. Depending on your require-
ments, you can extend or further restrict the tolerance
ranges in the Test menu under the menu item Check
limits (e.g. for low limit values for the bath control).