Use and Operation
6-6
adjust mA. When mA is adjusted, the power will not exceed the maximum output
power of the device, and the device will intelligently recognize the limit. APR1~
APR8 cannot be adjusted in mA adjustment mode. After mA adjustment is
completed, long press APR8 to return to display the exposure parameters of kV
and mAs.
f) Adjust the exposure parameters with
button.
g) Select the proper exposure parameter.
When using the exposure hand switch, press the hand switch Level 1, or
when using the remote control, press the "A" key, the remote control indicator
turns on and the X-ray machine makes a "Di" sound. Press the "B" key within 3
seconds, the remote control indicator turns on and the X-ray machine makes a
"Di" sound. When
Exposure Level 1 ready indicator is on, the device
enters Exposure Level 1 Ready state.
h) When using the exposure hand switch, press hand switch Level 2 or use the
remote control to do nothing after the Indicator
Exposure Level 1 is
on. When
exposure indicator is on, the device enters the exposure
state. When both
and indicators go off, one normal
exposure is accomplished and
turns on. Now the user can release
the hand switch.
i) Wait for 5s, then the user may repeat steps e) to h) and do another exposure.
j) Press
to turn off the device, and all the digital tube displays on the
control panel go off.
k) Toggle the rocker switch on the back cover of the device to "OFF".
【Caution】
and :
When
is on, it means the device is normal, without any