Use and Operation
6-13
exposure is triggered, the exposure countdown is entered, the exposure delay is 0 to 60
seconds, the default is 5 seconds, 0 is no delay, that is, the second press of the exposure
button will perform the exposure immediately, the exposure light is on; if it is not 0, the
exposure is triggered, the READY indicator flashes at a delay time of 3 times per second
till completion, and the exposure starts at the end of the READY time. After the exposure
is over, the exposure indicator turns off and the COMPLETE indicator turns on.
For the preparation time setup of one-press exposure, see P27 (Preparation time setting
of one-press exposure button) of 6.1.3 Engineering Mode Function and Operation.
For the Exposure delay setting of one-press exposure button, see P28 (Exposure delay
setting of one-press exposure button) of 6.1.3 Engineering Mode Function and
Operation.
Fig. 6-5
Caution
a)
Wait for 5s between two consecutive exposure operations;
otherwise normal exposure cannot be performed.
b) When the portable X-ray machine is turned on for a period of
time and if no operation is performed, it will enter the
automatic standby state. If it is not turned on for a period of
time in the standby state, the X-ray machine will automatically
shut down. See P6 (Standby time) and P21 (The timing setting
from shutdown to poweroff) of 6.1.3 Engineering Mode
Function and Operation.
c) There is an icon in the upper right of the portable X-ray
machine for power display. When the power is low, the device
shall be recharged in time. Make sure that the power switch on