b. Switch Adjustment.
If proper readings are not obtained during test or if
table does not properly return to level, check the
tighness of the table top section fasteners to make
sure the sections are secure and use the following
procedure to adjust the switches.
1. Apply table brakes and (using a level) level
the table top using the TRENDELENBURG and
LATERAL-TILT function buttons on the pendant
control.
2. Carefully loosen the allen bolts securing the
switch bracket and adjust the switches as needed.
See figure 5-31.
Figure 5-31. Micro-Switch Adjustment.
Figure 5-30. Kidney Lift Inhibit Switch
82306-529
PINK
ORANGE/WHITE
PINK
ORANGE/WHITE
(CN10)
CS-5
COM
NC
TEST AT PINS 12 & 24
CS-5 KIDNEY LIFT INHIBIT
Table Position
Back-Up < 45º
Back-Up > 45º
Switch Position
Open
Closed
Meter Reading
Infinity
0
When the Back Section is more than 45º above horizontal,
CS-5 will not allow Kidney Lift to operate.