42
The following conventions are used to describe the commands syntax for the 4320/4520. Braces ({ }) enclose
each parameter for a command string. Triangle brackets (< >) indicate that you must substitute a value for the
enclosed parameter. The Pipe ( | ) is used to separate different parameter options for a command. The
command and the parameter data must be separated with a space. All commands that end with a question
mark (?) are query commands and require an IEEE-488 read command to retrieve the data from the device's
output buffer.
Test Execution Commands
The following commands are used to control actual output voltage and current from the tester. Please
observe all safety precautions.
COMMAND
DESCRIPTION
TEST
Execute a Test
RESET
Abort a Test in Process or Reset Failures
SAO
Set Auto Offset
TEST
Starts the test sequence at the selected step loaded into memory (RAM).
RESET
Stop or abort a test. Also used to reset a latched failure condition.
SAO
Set the offset for the Ground bond test. The cables and any test fixture should be connected before executing
the command. This command will perform an actual test and all safety precautions should be observed when
using this command.