Installation Flowchart
VII
Installation Flowchart
Yes
Was the transmitter
configured on the bench
to match the application?
Configure the action and
the limit (Seção 3).alarm
Simule the value(s)
of the work range in the sensor(s)
conection terminal(s)
mV or Ohm
Start
No
No
OK
Yes
No
Yes
No
Yes
Configure the transmitter 1
and 3 - Configuration)
(Section
Section
Configure the measuring range
to 0% (4mA) and 100% (20mA)
(Section 3 - Configuration)
Configure the Damping
(Section 3 - Configuration)
Configure the LCD reading
(Section 3 - Configuration)
See manual
(Section 5 - ) Maintenance
Is the reading correct?
Check the area classification
and its practices.
Install the transmitter preferably
on weather- protected areas.
Install the transmitter
(mechanically and electrically)
according to the application after
checking the best position for
the LCD (Section 5 - Maintenance)
Power the transmitter properly.
Is the transmitter
reading correct?
Do Zero and Span Trim
Did you correct the
transmitter reading?

Table of Contents