Solartron 1287 - Error; Warning Code Summary

Default Icon
134 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1 ERROR
/
WARNING
CODE SUMMARY
1.1
GROUP 0 : Command Structure
Errors
01 Unknown command.
02
Argument mismatch.
03 Argument out of range.
04
Floating point format error.
05 Illegal request for value.
1.2 GROUP
1 : Store Errors
1 1 Store does not exist.
1
2
Store check
sum error.
1
3 Not enough memory for store.
1 4 Requested store is
not complete.
1
5 Linked list in store corrupted.
1.3 GROUP
2
: Parametric
Interaction
Errors
21
IRC not allowed in galvanostatic
(G STAT) mode.
22 Polarisation current
(POL I) and/or current
bias rejection
(I REJECT) inconsistent with standard
resistor selected.
23 Necessary parameter
in sweep notpntered.
24
Zero
or negative
values not allowed in
log plots
or
VTIME
plots.
25 VX MIN > VX
MAX
and/or VY
MIN > V
Y MAX in the PLOT SIZE
menu.
26 IRCQ or RPCO
inconsistent with standard
resistor selected.
27 GPIB
echo not allowed in plotting.
28 Ramp sweep: ramp
rate
>100V/s
Step sweep: V/STP<100pV
when maximum
excursion from RMP
Vl>200mV;
I/STP x standard
resistor <100pV when
maximum
excursion (x
standard resistor) from STP Il>200mV.
29 Ramp
sweep: ramp rate
dOOpV/s
Step sweep: V/STP
<50|iV when maximum excursion
from RMP
V1 <200mV
and >20mV;
I/STP x standard resistor <50jiV
when maximum excursion
from STP
11 <200mV and >20mV.
1.4
GROUP
3 : Overloads and
Warnings
3 1 Current DVM overload.
32 Voltage DVM overload.
33 Both DVMs
overload.
34 Thermal cut-out. Over-heating
of the amplifier
output stage has been
detected and the
instrument has switched
to standby. Over-heating
may
be caused by: a) excessive ambient
temperature
(>50C), b) cooling
fan failure or blocked vents
or c) prolonged high dissipation
e.g. battery
discharge testing.
The output cannot be switched
on again until
a
safe
temperature
is regained.
35 RE1 overload.
36 RE1-RE2 overload.
37 Current overload.
38 Current limit.
39 SI1 287 standby.
CSB/SI1287 User
Guide/Issue AA
B.3