MPI-520 OPERATING MANUAL version 3.4
38
Remarks:
ATTENTION!
When Object under voltage message is displayed, the object tested is live. The
measurement is blocked. The meter must be immediately disconnected from the
object.
Additional information displayed by the meter
NOISE!
Interference voltage occurs on the object tested. The
measurement is possible however it will be burdened with
additional uncertainty that is specified in the technical data.
3.8.2 Measurement of resistance
Set the rotary switch of function selec-
tion at R
X
R
±200 mA
position.
Press F1 push-button and move to selection of
measurement mode.