MPI-520 OPERATING MANUAL version 3.4
39
Select R
X
position by means of and
push-buttons and confirm by pressing
ENTER.
Connect the meter to the object tested.
N
L
Read out the result.
Remarks:
- Remarks and messages are the same as in point 3.8.1.
3.8.3 Calibration of test leads
In order to eliminate the impact of the resistance of test leads on measurement result, the com-
pensation (autozeroing) of resistance may be performed. For this purpose, R
X
and R
±200 mA
functions
have AUTOZERO sub-function.
Press F2 push-button.