S6 Portable Digital Color Doppler Ultrasound System
Service Manual
Operating Problems: continued from previous page
Problem
Symptoms
Suggested Solutions
System freezes
After being in Freeze mode
for several hours, the
system is not responding
to the FREEZE key, or any
other keyboard inputs.
Need to use the mains
POWER switch on the
rear panel to turn off the
system. The system seems
to be working normally
after restarting.
Update the software
Slow system re-
sponse.
Startup takes too much
time (a 30 minutes’ startup
has encountered by a cus-
tomer). The system is re-
sponding slowly. While
saving images is normal,
reading may take as long as
30 minutes.
Update the software. If problem persists, re-
place the HDD.
^
Shutdown Problems
Problem
Symptoms
Suggested Solutions
Unable to turn off
the system nor-
mally
The system won’t shut
down after the POWER
button on the side panel is
pressed.
Press the power button for a few seconds until
the system is powered off.
Reinstall the system if the same problem oc-
curs frequently.
Shutdown ends
up with a Linux
interface
After POWER button is
pressed, a Linux interface
appears instead of normal
shutdown window.
Press the power button for a few seconds until
the system is powered off.
Reinstall the system if the same problem oc-
curs frequently.
Freeze at shut-
down window
After POWER button is
pressed, the system freezes
at the shutting down win-
dow.
Press the power button for a few seconds until
the system is powered off.
Reinstall the system if the same problem oc-
curs frequently.
Unable to turn off
the system even
long-press the
POWER button
The system won’t shut
down even long pressing
the POWER button on the
side panel.
Use the mains power switch on the rear panel
to power off the equipment.
Reinstall the system and update the software.
Keyboard remains
illuminated
Unable to turn off the
system completely: after
POWER button is pressed,
the monitor turns dark but
the keyboard remains illu-
minated.
Replace the RAM.
Update the software.
P/N: 4720-0034-01A