S6 Portable Digital Color Doppler Ultrasound System
Service Manual
Startup Problems: continued from previous page
Problem
Symptoms
Suggested Solutions
Brown screen af-
ter system boot-
ing up - II
The screen turns brown,
displaying the following
messages after startup:
ID=0.2
Cmd Ack Timeout Exp=3,Res=1
Command IN Ack Timeout
EC Header Corrupted (255)!
Swap IN pingpong 1 . . .
Flag=0
Send our dummy packet...
Flag=1
wap IN pingpong 2. . .
Flag=0
Restart the system several times to check if
the same problem occurs.
If the problem does occur again, update the
software.
If the problem persists, it might be caused by
the DBF board malfunction.
Freezes at the
booting up screen
The system freezes at the
"System Booting" screen.
Update the software.
Slow start up
Blue screen appears after
the system is turned on.
The system is able to en-
ter the EXAM screen af-
ter restarting three times,
however, startup process
may take more than 10
minutes.
Update the software.
If the update fails, or the update does not
solve the problem, replace the HDD with an-
other one which has the updated software in-
stalled in.
P/N: 4720-0034-01A