Heater-Cooler System 3T • Safety
2.6 CP_IFU_16-XX-XX_USA_015
2.2.6 Safety instructions for routine maintenance
◗ Routine maintenance work must only be performed by qualified personnel.
◗ The heater-cooler must be cleaned completely prior to operating it for the first time.
◗ The water circuits must be disinfected prior to operating the heater-cooler for the first time and
when placing the system in storage. The disinfection cycle must then be repeated every 14 days
(even on systems in storage).
!
◗ Disconnect the heater-cooler from the power supply prior to performing routine maintenance.
◗ Do not fail to observe the regulations concerning routine maintenance, as well as the prescribed
maintenance intervals stated in these operating instructions.
◗ Ensure that no liquids or cleaning agents enter the housing through the vents or other openings.
◗ Use recommended cleaning agents and disinfectants. Follow the instructions of the chemical
manufacturer for storage and use condition.
◗ Repair work on the heater-cooler must only be carried out by authorized service personnel. Only
original spare parts from SORIN GROUP DEUTSCHLAND GMBH must be used in order to guarantee
the proper functioning of the system.
◗ The repair work of the cooling circuit and the disposal of the refrigerant must only be performed by
an expert for refrigerant systems. Please see chap. 6.1.3 “Regular maintenance checks by
authorized service technicians” on page 6.4.
◗ Routine decalcification is not required. If needed, decalcification must be carried out by
authorized service personnel.