Heater-Cooler System 3T • Appendix
7.2 CP_IFU_16-XX-XX_USA_015
7.1.3 General performance data
General data
Heating performance (2x pat., 1x c-plegia) 3 x 1,350 W (input voltage 240 V ~)
3 x 1,000 W (input voltage 208 V ~)
3 x 650 W (input voltage 120 V ~)
Cooling performance > 1,500 W at an ambient and tank
temperature of 20°C / 68°F
Noise level ≤ 60 dB (A)
Volume and flow
Minimum level patient circuits 4.5 l (11.6 l total volume)
Maximum level patient circuits 6.5 l (13.8 l total volume)
Water flow patient circuits 15 - 17 l
Water flow cardioplegia circuit 9 - 11 l
Water pressure in the circuits 0.7 bar max.
-0.3 bar min.
Temperature ranges
Normal range patient circuits from 2°C / 35.6°F through 41°C / 105.8°F
± 0.5°C / 0.9°F
Normal range cardioplegia-cooling circuit from 2°C / 35.6°F through 10°C /50°F
± 2°C / 3.6°F
Normal range cardioplegia-heating circuit from 15°C / 59.0°F through 41°C / 105.8°F
± 1°C / 1.8°F