Heater-Cooler System 3T • Routine maintenance
6.8 CP_IFU_16-XX-XX_USA_015
6.3.1 Disinfection procedure
Make sure that the heater-cooler is switched off.
Drain the water tanks.
Close the drain valves 11 and 12.
Make sure that the three venting valves at the rear of the heater-
cooler are closed.
Press the mains power switch 1 to power up the heater-cooler.
Press the key Audio alarm off 29.
Remove the cover of the filler neck 3 by turning it
90° counterclockwise.
Fill the water tanks with filtered tap water until the orange LED b
of the water level display for the patient circuit lights up.
Dose the disinfectant to be used (Minncare Cold Sterilant or
Clorox Regular Bleach) with a measuring cup:
450 ml (15 US fl. oz.) of concentrated Minncare Cold Sterilant
Note: Minncare Cold Sterilant mixed with filtered tap water at a 1:30
ratio results in a concentration of use dilution of 3.3%.
OR
180 ml (6 US fl. oz.) of concentrated Clorox Regular Bleach
Note: Clorox Regular Bleach solution (active ingredient: sodium
hypochlorite concentration 8.25%) mixed with filtered tap water at
a 1:77 ratio results in a concentration of use dilution of 1.3%.
Pour the disinfectant into the water tanks.
Complete filling the tanks with filtered tap water until the second
green LED d of the water level display for the patient circuit lights up.