Heater-Cooler System 3T • Appendix
CP_IFU_16-XX-XX_USA_015 7.7
7.6 Information about electromagnetic
compatibility (EMC) for 240 V ~ and 208 V ~
System
7.6.1 Guidance and manufacturer's declaration
Note:
!
Medical electrical equipment needs precautions regarding electromagnetic compatibility and
has to be installed and put into service according to the EMC information provided in the
following guidance and the manufacturer's declaration.
!
Portable and mobile RF-communications equipment can affect medical electrical equipment.
Guidance and manufacturer's declaration – electromagnetic emission
The Stöckert Heater-Cooler System 3T is intended for use in the electromagnetic environment specified below. The customer or
the user of the Stöckert Heater-Cooler System 3T should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment - guidance
RF emissions CISPR 11 Group 1 The Stöckert Heater-Cooler System 3T uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
RF emissions CISPR 11 Class B The Stöckert Heater-Cooler System 3T is suitable for use in all
establishments, including domestic establishments and those directly
connected to the public low-voltage power supply network that supplies
buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations /
flicker emissions IEC 61000-3-3
and IEC 61000-3-2
Complies However, the device must only be connected to a power supply network
the impedance of which is less than or equal to 0.391Ω + j0.244Ω .
a
a
NOTE: Only for Heater-Cooler System 3T 240 V ~ and 208 V ~ .